您好, 歡迎來到化工儀器網(wǎng)! 登錄| 免費注冊| 產(chǎn)品展廳| 收藏商鋪|
當(dāng)前位置:上海境宣實業(yè)有限公司>>日本光明北川式氣體檢測管>>氣體快速檢測管>> 157SB甲基溴氣體檢測管157SB METHYL BROMIDE 20-250ppm
參 考 價 | 面議 |
產(chǎn)品型號157SB
品 牌
廠商性質(zhì)代理商
所 在 地上海市
更新時間:2024-11-20 10:29:31瀏覽次數(shù):1646次
聯(lián)系我時,請告知來自 化工儀器網(wǎng)室內(nèi)空氣甲醛檢測管No.713 FORMALDEHYDE 0.01-0
氯乙烯CH2=CHCl氣體測毒管132SC VINYL CHLORID
氯乙烯CH2=CHCl氣體檢知管132SB VINYL CHLORID
甲基溴氣體檢測管157SB METHYL BROMIDE 20-250ppm
157SB
1. PERFORMANCE
1) Measuring range :10-500ppm
Number of pump strokes 1(100ml)
2) Sampling time :2 minutes/1 pump stroke
3) Detectable limit :1 ppm
4) Shelf life :3 years(Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :White→Reddish orange
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
By decomposing with an Oxidizer, Bromine is produced. Bromine reacts with o-Toluidine and red Orthoquinone
is produced.
CH3Br+I2O5+CrO3+H2SO4→Br2
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
10
30
50
100
150
200
300
400
500
ppm
100ml
GAS To pump
(Brown) (White)
PRETREAT TUBE DETECTOR TUBE
Tube No.
157SA METHYL BROMIDE
Substance ppm Interference ppm Coexistence
Chlorine 1 Similar stain is produced. 1 Higher readings are given.
Bromine 1 〃1 〃
Nitrogen dioxide 1 〃1 〃
Dichloromethane 500 〃500 〃
Chloroform 50 〃50 〃
Ethylene dibromide 50 〃50 〃
Trichloroethylene 50 〃50 〃
Tetrachloroethylene 50 〃50 〃
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 0℃ 5℃ 10℃ 15℃ 20℃ 30℃ 40℃
(ppm) (32°F) (41°F) (50°F) (59°F) (68°F) (86°F) (104°F)
500 - - - 620 500 430 390
400 - - 630 475 400 360 325
300 - 675 430 345 300 280 260
200 580 320 235 210 200 190 175
150 260 185 160 150 150 140 130
100 120 110 100 100 100 95 90
50 50 50 50 50 50 50 50
185
1. PERFORMANCE
1) Measuring range :5-80 ppm 2.5-40 ppm 0.9-7.3 ppm
Number of pump strokes 1(100ml) 2(200ml) 4(400ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :1 ppm(200ml)
4) Shelf life :3 years(Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :White→Yellow
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
By decomposing with an Oxidizer, Bromine is produced. Bromine reacts with o-Toluidine and yellow
Orthoquinone is produced.
CH3Br+I2O5+CrO3+H2SO4→Br2
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
5
10
15
20
30
40
60
80
ppm
100ml
GAS To pump
(Olive) (White)
PRETREAT TUBE DETECTOR TUBE
Tube No.
157SB METHYL BROMIDE
(NOTE)
When the concentration is below 5 ppm, 2 pump strokes can be used to determine the lower concentration.
Following formula is available for the actual concentration.
Actual concentration=1/2×Temperature corrected value
Substance Interference ppm Coexistence
Halogen Similar stain is produced. Higher readings are given.
Halogenated hydrocarbons 〃〃
Hexane FIG.1 200 Lower readings are given.
0
50
60
70
80
30
40
Hexane〔ppm〕?
FIG.1 Influence of Hexane
請輸入賬號
請輸入密碼
請輸驗證碼
以上信息由企業(yè)自行提供,信息內(nèi)容的真實性、準(zhǔn)確性和合法性由相關(guān)企業(yè)負(fù)責(zé),化工儀器網(wǎng)對此不承擔(dān)任何保證責(zé)任。
溫馨提示:為規(guī)避購買風(fēng)險,建議您在購買產(chǎn)品前務(wù)必確認(rèn)供應(yīng)商資質(zhì)及產(chǎn)品質(zhì)量。