當前位置:上海境宣實業(yè)有限公司>>日本光明北川式氣體檢測管>>氣體快速檢測管>> 157SB溴甲烷CH2Br2氣體檢測管157SB METHYL BROMIDE 2.5-40ppm
室內空氣甲醛檢測管No.713 FORMALDEHYDE 0.01-0
氯乙烯CH2=CHCl氣體測毒管132SC VINYL CHLORID
氯乙烯CH2=CHCl氣體檢知管132SB VINYL CHLORID
溴甲烷CH2Br2氣體檢測管157SB METHYL BROMIDE 2.5-40ppm
1. PERFORMANCE
1) Measuring range :5-80 ppm 2.5-40 ppm 0.9-7.3 ppm
Number of pump strokes 1(100ml) 2(200mr) 4(400ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :1 ppm(200ml)
4) Shelf life :3 years(Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :White→Yellow
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
By decomposing with an Oxidizer, Bromine is produced. Bromine reacts with o-Toluidine and yellow
Orthoquinone is produced.
CH3Br+I2O5+CrO3+H2SO4→Br2
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
5
10
15
20
30
40
60
80
ppm
100ml
GAS To pump
(Olive) (White)
PRETREAT TUBE DETECTOR TUBE
Tube No.
157SB METHYL BROMIDE
(NOTE)
When the concentration is below 5 ppm, 2 pump strokes can be used to determine the lower concentration.
Following formula is available for the actual concentration.
Actual concentration=1/2×Temperature corrected value
Substance Interference ppm Coexistence
Halogen Similar stain is produced. Higher readings are given.
Halogenated hydrocarbons 〃〃
Hexane FIG.1 200 Lower readings are given.
0
50
60
70
80
30
40
Hexane〔ppm〕?
FIG.1 Influence of Hexane