產(chǎn)品簡介
詳細介紹
For the Fast Isocratic Separation of Inorganic Anions and Oxyhalides
The IonPac® AS9-SC is designed for the isocratic separation of oxyhalides and inorganic anions, including fluoride, chlorite, bromate, chloride, nitrite, bromide, chlorate, nitrate, phosphate, and sulfate in drinking water, wastewater, groundwater, and other diverse sample matrices.
For fast analysis of oxyhalides and inorganic anions at similar concentrations
Ideal for simple sample matrices
Common inorganic anions and oxyhalides can be determined in less than 10 minutes using an isocratic carbonate/bicarbonate eluent coupled with suppressed conductivity detection.
The IonPac AS9-SC column is specified in US EPA Method 300.0 (B), and meets or exceeds the performance requirements of this method. Use the Anion Self-Regenerating Suppressor (ASRS® 300) with the IonPac AS9-SC column.
Note: See also the IonPac AS23, the recommended column for the analysis of oxyhalides, including bromate.
IonPac AS9-SC Anion-Exchange Column Specifications | |
---|---|
Dimensions | IonPac® AS9-SC Analytical Column: 4 × 250 mm IonPac AG9-SC Guard Column: 4 × 50 mm |
Maximum operating pressure | 4000 psi |
phase compatibility | pH 2–11; 0–* HPLC solvents |
Substrate characteristics | Bead Diameter: 13.0 µm Cross-Linking (%DVB): 55% |
Latex characteristics | Functional Group: Alkyl quaternary ammonium ion Latex Cross-Linking: 20% Latex Diameter: 110 nm Hydrophobicity: Medium |
Capacity | 30–35 µeq (4 × 250 mm analytical column) |
Column construction | PEEK™ with 10-32 threaded ferrule-style end fittings. All components are nonmetallic. |