800臺式電動離心機具有造型美觀,容量大,體積小、功能齊全。其性能穩(wěn)定,速度可調(diào)并能自動調(diào)節(jié)平衡,溫升低使用效率高,以及適用性廣等特點。該產(chǎn)品適用于醫(yī)藥制品、血站、臨床試驗及生物化學(xué)實驗室作血清、血漿、尿素的定性分析。
800臺式電動離心機
800型臺式電動離心機具有造型美觀,容量大,體積小、功能齊全。其性能穩(wěn)定,速度可調(diào)并能自動調(diào)節(jié)平衡,溫升低使用效率高,以及適用性廣等特點。
該產(chǎn)品適用于醫(yī)藥制品、血站、臨床試驗及生物化學(xué)實驗室作血清、血漿、尿素的定性分析。
一、800臺式電動離心機技術(shù)參數(shù)
型號:800電動離心機
調(diào)速范圍(rpm):0~4000
容量(ml):20×6
離心力(xg):1430
電源:220V 50HZ
800D Desk centrifugal settler
Based on the 800 settler, this centrifugal settler belonging to advanced generation with many advantages including beautiful appearance、small volume、light weight、high capacity, so it applies to the qualitative analysis of plasma and serum in hospital labouratory and biochemistry laboratory as well as the carbamide and vaccine processing.
Technical parameters:
1,Power supply: 110V 50Hz
2, Timing range:0-60min
3, Rated power:50W
4: Speed range:0-4000 r/min
5, Capacity: 20ml X6(Angie type)
6, Separating factor:1430
Important Notice:
For your safety, a reliable grounding is strongly recommended to be done before operation,
Before the first time use, please read this Operation Manual carefully.
Working conditions: 0-40℃ambient temperature and 80% below relative humidity.
It is suggested to cut off power when long term nonuse.
Operation Procedure:
Insert the plug into socket and turn on the power switch.
Put the medicine into centrifuge tube and then insert the tube into the corresponding sleeve, close the cover plate and select proper centrifugal speed according to test requirements, then select running time or close ON. So it is easy to operate this instrument with better effect.
Daily maintenance
Correct operation and maintenance can extend the service life, it is suggested to place this instrument under reasonable working conditions so that the instrument can work normally.
If the liquid is splashed or sprayed into instrument by accident, cut off the power supply and remove the residual liquid in case shorten the service life.
One year warranty and lifelong maintenance are offered to our product (reasonable labor and material cost will be triggered by man-made damage or misoperation
二、800電動離心機安全須知
嚴禁在未裝轉(zhuǎn)頭的情況下空載運行;運行之前應(yīng)檢查轉(zhuǎn)頭有無腐蝕損傷;當(dāng)轉(zhuǎn)頭運行時不要接觸正在運行的轉(zhuǎn)頭;嚴禁轉(zhuǎn)頭在裝載不平衡的狀態(tài)下運轉(zhuǎn)。
三、800電動離心機使用說明
離心機的工作臺應(yīng)平整堅固,工作間應(yīng)整齊清潔,干燥并通風(fēng)良好,環(huán)境溫度以5-32°C為宜.開啟離心機將內(nèi)腔及轉(zhuǎn)頭擦拭干凈.裝放稱量*的試管,關(guān)閉離心機蓋,調(diào)節(jié)調(diào)速旋鈕,升至所需轉(zhuǎn)速.每次停機前,必須將調(diào)速旋鈕置于zui小位置.“定時器”置“零”關(guān)電源開關(guān),切斷電源.擦拭內(nèi)腔及轉(zhuǎn)頭,關(guān)閉離心機蓋。